Invento - NC Post Dilatation Catheter

The Invento NC Rx is a High Pressure Capacity and controlled expansion of non-compliant Ballon. It's a true Noncompliant Ballon with an expansion of 5%

TECHNICAL PARAMETERS

Design
Laser Cut Closed Cell
Non Compliant polyamide Ballon
Precise & Predictable ballon growth
Hydrophilic coating
On the distant Shaft
Radio Opaque markers
Zero Profile
Maximum recommended guidewire
0.014″ (0.36mm)
Excellent deflation time
Average less than <_15 sec
Proximal Shaft diameter
Proximal 1.19 F (0.73mm)
Distant Shaft Diameter
Distal 2.64 F(0.89mm)
Usable Catheter Length
1414 mm
Rapid Exchange Length
250 mm
Tip Entry Profile
0.016″(0.40 mm)

COMPLIANCE CHART

Pressure Atm
2.00
2.50
2.75
3.00
3.50
4.00
6
1.94
2.41
2.68
2.93
3.41
3.90
8
1.97
2.45
2.70
2.96
3.45
3.95
10
2.00
2.50
2.75
3.00
3.50
4.00
12
2.03
2.54
2.79
3.05
3.55
4.05
14
2.06
2.57
2.83
3.10
3.60
4.11
16
2.09
2.61
2.87
3.15
3.65
4.16
18
2.12
2.65
2.92
3.20
3.69
4.20
20
2.16
2.68
2.97
3.25
3.73
4.25
22
2.20
2.72
3.01
3.29
3.78
4.29

Rated Burst Pressure - Minimum Burst Pressure 22 ATM - Ballon size at 37.5 degree C - Tolerance +/- 0.10mm between 8 and 16 ATM.

ORDERING INFORMATION

Length (mm)
2.00
2.50
2.75
3.00
3.50
4.00
10 mm
INV NC 200/10
INV NC 250/10
INV NC 275/10
INV NC 300/10
INV NC 350/10
INV NC 400/10
12 mm
INV NC 200/12
INV NC 250/12
INV NC 275/12
INV NC 300/12
INV NC 350/12
INV NC 400/12
15 mm
INV NC 200/15
INV NC 250/15
INV NC 275/15
INV NC 300/15
INV NC 350/15
INV NC 400/15