TECHNICAL PARAMETERS

Design
Laser Cut Closed Cell
Semi Compliant polyamide Ballon
Precise & Predictable ballon growth
Hydrophilic coating
On the distant Shaft
Radio Opaque markers
Zero Profile
Maximum recommended guidewire
0.014″ (0.36mm)
Maximum Guiding Catheter Compatibility
5F
Average Recoil
<5%
Excellent deflation time
Average less than 10 sec
Proximal Shaft diameter
1.19 F (0.73mm)
Useable Catheter length
1400 mm
Rapid Exchange Length
250 mm
Radio Opaque Visual Marker
90 cm and 100 cm
Tip Entry Profile
0.016″(0.40 mm)

COMPLIANCE CHART

Pressure Atm
1.25
1.50
2.00
2.25
2.50
2.75
3.00
3.50
4.00
2
0.96
1.21
1.71
1.96
2.21
2.46
2.71
3.21
3.71
4
1.06
1.31
1.81
2.06
2.31
2.56
2.81
3.31
3.81
6
1.16
1.41
1.91
2.16
2.41
2.66
2.91
3.41
3.91
8
1.25
1.50
2.00
2.25
2.50
2.75
3.00
3.50
4.00
10
1.29
1.54
2.06
2.31
2.56
2.81
3.06
3.56
4.08
12
1.33
1.58
2.12
2.37
2.62
2.87
3.12
3.62
4.16
14
1.37
1.62
2.18
2.43
2.68
2.93
3.18
3.68
4.24
16
1.41
1.66
2.24
2.49
2.74
2.99
3.24
3.74
4.30
18
1.44
1.69
2.27
2.52
2.78
3.03
3.29
3.80
4.36
20
1.47
1.72
2.30
2.55
2.82
3.07
3.34
3.86
4.42

Rated Burst Pressure - Minimum Burst Pressure 22 ATM - Ballon size at 37.5 degree C - Tolerance +/- 0.10mm between 8 and 16 ATM.

ORDERING INFORMATION

Length (mm)
1.25
1.50
2.00
2.50
2.75
3.00
3.50
4.00
10 mm
INV NC 125/10
INV NC 150/10
INV NC 200/10
INV NC 250/10
INV NC 275/10
INV NC 300/10
INV NC 350/10
INV NC 400/10
12 mm
INV NC 125/12
INV NC 150/12
INV NC 200/12
INV NC 250/12
INV NC 275/12
INV NC 300/12
INV NC 350/12
INV NC 400/12
15 mm
INV NC 125/15
INV NC 150/15
INV NC 200/15
INV NC 250/15
INV NC 275/15
NINV NC 300/15
INV NC 350/15
INV NC 400/15
Translate »