Pressure Atm | 2.25 | 2.50 | 2.75 | 3.00 | 3.50 | 4.00 |
---|---|---|---|---|---|---|
2 | 1.96 | 2.21 | 2.46 | 2.71 | 3.21 | 3.71 |
4 | 2.06 | 2.31 | 2.56 | 2.81 | 3.31 | 3.81 |
6 | 2.16 | 2.41 | 2.66 | 2.91 | 3.41 | 3.91 |
8 | 2.25 | 2.50 | 2.75 | 3.00 | 3.50 | 4.00 |
10 | 2.31 | 2.56 | 2.81 | 3.06 | 3.56 | 4.08 |
12 | 2.37 | 2.62 | 2.87 | 3.12 | 3.62 | 4.16 |
14 | 2.43 | 2.68 | 2.93 | 3.18 | 3.68 | 4.24 |
16 | 2.49 | 2.74 | 2.99 | 3.24 | 3.74 | 4.30 |
18 | 2.52 | 2.78 | 3.03 | 3.29 | 3.80 | 4.36 |
20 | 2.55 | 2.82 | 3.07 | 3.34 | 3.86 | 4.42 |
Design | Laser Cut Closed Cell |
---|---|
Stent Material | MP35N Cobalt Chromium Alloy |
Stent Platform | Co-Cr Stent |
Strut Thickness / Width | 0.80 mm / 0.075 mm (SV)
0.092 mm / 0.083 mm (MV)
|
Metal to Artery Ratio | 13.9 – 19.1% (SV)
11.4-15.2%(MV) |
Average Foreshortening | < 2% |
Average Recoil | <5% |
Average Crossing Profile | <1.0 mm |
Drug | Sirolimus (Rapmycin) |
Polymers Type | Fully Biodegradable Polymer |
SV – Small Vessel Stent
MV – Medium Vessel Stent | Total Release of Drug in 45 – 48 days
Total loading 161 mg in 3.0 * 19 mm Stent |
Length (mm) | 2.25 | 2.50 | 2.75 | 3.00 | 3.50 | 4.00 |
---|---|---|---|---|---|---|
11 mm | NVC225/11 | NVC250/11 | NVC275/11 | NVC300/11 | NVC350/11 | NVC400/11 |
14 mm
| NVC225/14 | NVC250/14 | NVC275/14 | NVC300/14 | NVC350/14 | NVC400/14 |
19 mm | NVC225/19 | NVC250/19 | NVC275/19 | NVC300/19 | NVC350/19 | NVC400/19 |
24 mm | NVC225/24 | NVC250/24 | NVC275/24 | NVC300/24 | NVC350/24 | NVC400/24 |
29 mm | NVC225/29 | NVC250/29 | NVC275/29 | NVC300/29 | NVC350/29 | NVC400/29 |
34 mm | NVC300/34 | NVC350/34 | NVC400/34 | |||
39 mm | NVC300/39 | NVC350/39 | NVC400/39 |